{"_buckets": {"deposit": "6e66efca-d0c6-4351-b146-3326f15a4806"}, "_deposit": {"created_by": 1, "id": "32590", "owners": [1], "pid": {"revision_id": 0, "type": "depid", "value": "32590"}, "status": "published"}, "_oai": {"id": "oai:jaxa.repo.nii.ac.jp:00032590", "sets": ["1890", "1910"]}, "author_link": ["410250", "410249"], "item_9_alternative_title_2": {"attribute_name": "その他のタイトル(英)", "attribute_value_mlt": [{"subitem_alternative_title": "A Cross-Spectrum Analyzer"}]}, "item_9_biblio_info_10": {"attribute_name": "書誌情報", "attribute_value_mlt": [{"bibliographicIssueDates": {"bibliographicIssueDate": "1980-11", "bibliographicIssueDateType": "Issued"}, "bibliographicIssueNumber": "4_A", "bibliographicPageEnd": "1282", "bibliographicPageStart": "1267", "bibliographicVolumeNumber": "16", "bibliographic_titles": [{"bibliographic_title": "東京大学宇宙航空研究所報告"}]}]}, "item_9_description_16": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "この論文は,分析フィルタとして低域フィルタを用いた,可聴周波数範囲の信号のアナログ型クロススペクトル分析計について,その動作原理,試作装置,測定結果などを報告するものである.分析計の周波数分解能は使用する低域フィルタの帯減幅によって決定される.試作した装置ではこの分解能を0.3Hzまで小さくすることが可能であるが,この値は分析フィルタとして帯減フィルタを用いた従来のクロススペクトル分析計では到達しえないものである.また分析フィルタの位相特性をそろえるという点についても,低減フィルタの使用は,その時定数を合わせるだけですむので,帯減フィルタを用いる場合にくらべてきわめて有利である.", "subitem_description_type": "Abstract"}]}, "item_9_description_17": {"attribute_name": "抄録(英)", "attribute_value_mlt": [{"subitem_description": "An analog cross-spectrum analyzer for audio frequency range using low-pass analyzing filters is described. The frequency resolution of the analyzer is equal to the band width of the low-pass filters and can be made as small as 0.3Hz, which is not attainable by the ordinary cross-spectrum analyzers using band-pass analyzing filters. The alignment of phase characteristics of the analyzing filters directly affects the accuracy of the measured spectra and is difficult for the band-pass filters. The use of low-pass filters as the analyzing filters also makes it easier because the only parameter to be adjusted is the time constants of the low-pass filters.", "subitem_description_type": "Other"}]}, "item_9_description_32": {"attribute_name": "資料番号", "attribute_value_mlt": [{"subitem_description": "資料番号: SA0125916000", "subitem_description_type": "Other"}]}, "item_9_publisher_8": {"attribute_name": "出版者", "attribute_value_mlt": [{"subitem_publisher": "東京大学宇宙航空研究所"}]}, "item_9_source_id_21": {"attribute_name": "ISSN", "attribute_value_mlt": [{"subitem_source_identifier": "0563-8100", "subitem_source_identifier_type": "ISSN"}]}, "item_9_source_id_24": {"attribute_name": "書誌レコードID", "attribute_value_mlt": [{"subitem_source_identifier": "AN00161914", "subitem_source_identifier_type": "NCID"}]}, "item_9_text_34": {"attribute_name": "メタデータ提供者", "attribute_value_mlt": [{"subitem_text_value": "メタデータ提供者: ISASNII"}]}, "item_9_text_35": {"attribute_name": "JAXAカテゴリ", "attribute_value_mlt": [{"subitem_text_value": "JAXAカテゴリ: 宇宙科学研究所"}]}, "item_9_text_43": {"attribute_name": "DSpaceコレクション番号", "attribute_value_mlt": [{"subitem_text_value": "DSpaceコレクション番号: 6"}]}, "item_9_text_6": {"attribute_name": "著者所属", "attribute_value_mlt": [{"subitem_text_value": "東京大学宇宙航空研究所"}]}, "item_9_text_7": {"attribute_name": "著者所属(英)", "attribute_value_mlt": [{"subitem_text_language": "en", "subitem_text_value": "Institute of Space and Aeronautical Science, University of Tokyo"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "石井, 泰"}], "nameIdentifiers": [{"nameIdentifier": "410249", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "ISHII, Yasushi", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "410250", "nameIdentifierScheme": "WEKO"}]}]}, "item_files": {"attribute_name": "ファイル情報", "attribute_type": "file", "attribute_value_mlt": [{"accessrole": "open_date", "date": [{"dateType": "Available", "dateValue": "2020-01-24"}], "displaytype": "detail", "download_preview_message": "", "file_order": 0, "filename": "SA0125916.pdf", "filesize": [{"value": "778.4 kB"}], "format": "application/pdf", "future_date_message": "", "is_thumbnail": false, "licensetype": "license_free", "mimetype": "application/pdf", "size": 778400.0, "url": {"label": "SA0125916.pdf", "url": "https://jaxa.repo.nii.ac.jp/record/32590/files/SA0125916.pdf"}, "version_id": "9ddee594-91a2-4881-9dba-f0de60526049"}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "jpn"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "departmental bulletin paper", "resourceuri": "http://purl.org/coar/resource_type/c_6501"}]}, "item_title": "\u003c論文\u003eクロススペクトル分析計", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "\u003c論文\u003eクロススペクトル分析計"}]}, "item_type_id": "9", "owner": "1", "path": ["1890", "1910"], "permalink_uri": "https://jaxa.repo.nii.ac.jp/records/32590", "pubdate": {"attribute_name": "公開日", "attribute_value": "2015-03-26"}, "publish_date": "2015-03-26", "publish_status": "0", "recid": "32590", "relation": {}, "relation_version_is_last": true, "title": ["\u003c論文\u003eクロススペクトル分析計"], "weko_shared_id": -1}
<論文>クロススペクトル分析計
https://jaxa.repo.nii.ac.jp/records/32590
https://jaxa.repo.nii.ac.jp/records/32590f2c4a23f-ab50-404d-8cd5-08e3f311d143
名前 / ファイル | ライセンス | アクション |
---|---|---|
SA0125916.pdf (778.4 kB)
|
|
Item type | 紀要論文 / Departmental Bulletin Paper(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2015-03-26 | |||||
タイトル | ||||||
タイトル | <論文>クロススペクトル分析計 | |||||
言語 | ||||||
言語 | jpn | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | departmental bulletin paper | |||||
その他のタイトル(英) | ||||||
その他のタイトル | A Cross-Spectrum Analyzer | |||||
著者 |
石井, 泰
× 石井, 泰× ISHII, Yasushi |
|||||
著者所属 | ||||||
東京大学宇宙航空研究所 | ||||||
著者所属(英) | ||||||
en | ||||||
Institute of Space and Aeronautical Science, University of Tokyo | ||||||
出版者 | ||||||
出版者 | 東京大学宇宙航空研究所 | |||||
書誌情報 |
東京大学宇宙航空研究所報告 巻 16, 号 4_A, p. 1267-1282, 発行日 1980-11 |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | この論文は,分析フィルタとして低域フィルタを用いた,可聴周波数範囲の信号のアナログ型クロススペクトル分析計について,その動作原理,試作装置,測定結果などを報告するものである.分析計の周波数分解能は使用する低域フィルタの帯減幅によって決定される.試作した装置ではこの分解能を0.3Hzまで小さくすることが可能であるが,この値は分析フィルタとして帯減フィルタを用いた従来のクロススペクトル分析計では到達しえないものである.また分析フィルタの位相特性をそろえるという点についても,低減フィルタの使用は,その時定数を合わせるだけですむので,帯減フィルタを用いる場合にくらべてきわめて有利である. | |||||
抄録(英) | ||||||
内容記述タイプ | Other | |||||
内容記述 | An analog cross-spectrum analyzer for audio frequency range using low-pass analyzing filters is described. The frequency resolution of the analyzer is equal to the band width of the low-pass filters and can be made as small as 0.3Hz, which is not attainable by the ordinary cross-spectrum analyzers using band-pass analyzing filters. The alignment of phase characteristics of the analyzing filters directly affects the accuracy of the measured spectra and is difficult for the band-pass filters. The use of low-pass filters as the analyzing filters also makes it easier because the only parameter to be adjusted is the time constants of the low-pass filters. | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 0563-8100 | |||||
書誌レコードID | ||||||
収録物識別子タイプ | NCID | |||||
収録物識別子 | AN00161914 | |||||
資料番号 | ||||||
内容記述タイプ | Other | |||||
内容記述 | 資料番号: SA0125916000 |